Isac Congress Digital Data Correct Mario Roederer ..I feel obliged to

by Mitch Haynes <mitchhaynes@[EMAIL PROTECTED] > Jun 15, 2008 at 06:51 PM

Re: Re digital data
This message: [ Message body ] [ More options ]
Related messages: [ Next message ] [ Previous message ] [ In reply
to ]
From: WEHICytometry <facs_copy@[EMAIL PROTECTED]
>
Date: Tue Mar 14 2006 - 23:28:16 EST

On 14/03/2006, at 9:03 AM, Mario Roederer wrote:

> Actually, there aren't several transforms -- basically there are
> only two ("Logicle", from Dave Parks and Wayne Moore at Stanford,
> used by DiVa and FlowJo; and the "HyperLog" from Bruce Bagwell at
> Verity used by WinList).  And, in fact, these two are
> mathematically nearly identical; the Logicle function is a slightly
> more complex, and slightly smoother version, but probably without
> noticeable visual difference when the same parameters to the
> function are used.  The primary difference between the two is that
> the Logicle algorithm provides for a mechanism to automatically
> select parameters to the function that are optimized to the actual
> distribution of the data--hence, the transformation can, if
> desired, be stronger or weaker for one fluorescence channel than
> for another (which is reasonable, as the magnitude of the error in
> the measurement is very different in these channels, and the error
> in the measurement is the primary reason we need the transforms!).
>

I feel obliged to correct Mario there; there are indeed more than
two.  Another is the "split scale" form that is used in the WEASEL
fcm analysis and display program (see http://www.wehi.edu.au/
cytometry/WEASELv2.html).  This transform is mathematically simpler
but, I assert, no less valid.  It has been used in WEASEL for a year
or so and was described to the Australasian Flow Cytometry Group
last
year (see the abstract at http://www.wehi.edu.au/cytometry/Abstracts/
AFCG05B.html).

Frank.

     |	  |  << The Cytometry Laboratory
      \__/ <<<< The Walter & Eliza Hall Institute
------!!<<<<<< 1G Royal Parade, Parkville
      /!!\ <<<< Victoria 3050, Australia
     o !! \  << ph: 61_3_9345 2540, fax: 61_3_9347 0852
Received on Wed Mar 15 16:18:00 2006
This message: [ Message body ]
Next message: Matt Gordon: "Re: Fixation of cells for TUNEL assay"
Previous message: Carol Oxford: "LSRII pressurized waste tank"
In reply to: Mario Roederer: "Re: Re digital data"
Contem****ary messages sorted: [ By Date ] [ By Thread ] [ By Subject ]
[ By Author ] [ By messages with attachments ]
This archive was generated by hypermail 2.1.8 : Thu Mar 16 2006 -
03:12:01 EST
 




 1 Posts in Topic:
Isac Congress Digital Data Correct Mario Roederer ..I feel oblig
Mitch Haynes <mitchhay  2008-06-15 18:51:22 

Post A Reply:
  Go here to Signup

AddThis Feed Button


About - Advertising - Contact - Frequently Asked Questions - Privacy Policy - Terms of Use - Signup

Contact
tan12V112 Tue Dec 2 14:55:18 CST 2008.